Instrumentation · Engineering · Package · PLC
0755-8222 1901 info@ainstru.com

质量流量计 8E3B

质量流量计  8E3B

质量流量计 8E3B

The robust Promass E has a long-standing reputation as a reliable solution accurately measuring liquids and gases in a wide range of standard applications in various industries. With its compact transmitter Promass E 300 offers high flexibility in terms of operation and system integration: access from one side, remote display, improved connectivity options. Heartbeat Technology ensures measurement reliability and enables extension of recalibration cycles.

The Promass E is robust and is an economical solution for long-term use in high-precision liquid and gas measurement in various standard applications across different industries. Combined with the compact transmitter, the Promass E 300 offers flexible operation and easy system integration: wiring on the same side of the instrument, remote display and improved connection options. Heartbeat Technology ensures compliance at all times and guarantees process safety.

Benefits
Economical: versatile instrument; preferred alternative to conventional volumetric flowmeters
Fewer process measuring points: multivariable measurement (flow, density, temperature)
Space-saving installation: no inlet/outlet straight run requirement
Complete view of process and diagnostic information: various user-configurable input/output combinations and fieldbuses
Reduced complexity and diversity: user-configurable input/output functions
Integrated verification: built-in Heartbeat Technology

Applications
The measuring principle is unaffected by the physical properties of the fluid (e.g. viscosity and density).
High-precision measurement of liquids and gases in standard applications, with broad applicability.

Device features
Compact dual measuring tube sensor
Medium temperature: max. +150 °C (+302 °F)
Process pressure: max. 100 bar (1450 psi)
Compact dual-compartment housing with up to 3 inputs/outputs
Backlit display, touch key operation and WLAN access
Remote display unit available

 

E+H Product Model Configuration Tool

Please enter the product model to configure

Product comparison

Please enter the product models to compare in the fields below; up to three products can be compared

 

   

Technical documentation

   

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

   

Measurement principle video

Source: Endress+Hauserpublic information from the website

Compare

Density

Measuring Principle Coriolis
Product Headline Flowmeter with minimized total cost of ownership and a compact, easily accessible transmitter.
Accurate measurement of liquids and gases for a wide range of standard applications.
Sensor features Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Compact dual-tube sensor. Medium temperature up to +150 °C (+302 °F).
Transmitter features Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Process pressure up to 100 bar (1450 psi). Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access.

Density/Concentration

Measuring Principle Coriolis
Product headline Flowmeter with minimized total cost of ownership and a compact, easily accessible transmitter.
Accurate measurement of liquids and gases for a wide range of standard applications.
Sensor features Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Compact dual-tube sensor. Medium temperature up to +150 °C (+302 °F).
Transmitter features Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Process pressure up to 100 bar (1450 psi). Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access.
Nominal diameter range DN 8 to 80 (⅜ to 3")
Wetted materials Measuring tube: 1.4539 (904L)
Connection: 1.4404 (316/316L)
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
Max. measurement error Mass flow (liquid): ±0.15 % (standard), ±0.10 % (option)
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.50 %
Density (liquid): ±0.0005 g/cm³
Measuring range 0 to 180 000 kg/h (0 to 6615 lb/min)
Max. process pressure PN 100, Class 600, 63K
Medium temperature range –40 to +150 °C (–40 to +302 °F)
Ambient temperature range Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F)
Sensor housing material 1.4301 (304), corrosion resistant
Transmitter housing material AlSi10Mg, coated; stainless steel for hygenic transmitter design
Degree of protection Standard: IP66/67, Type 4X enclosure
IP69
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available"
Outputs 3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output
Inputs Status input
4-20 mA input
Digital communication HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, MODBUS TCP over Ethernet-APL, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Power supply DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvals ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
Product safety CE, C-tick, EAC marking
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Marine approvals and certificates LR approval, DNV approval, ABS approval, BV approval, CCS approval
Pressure approvals and certificates PED, CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, EHEDG, cGMP

Gas

Measuring Principle Coriolis
Product headline Flowmeter with minimized total cost of ownership and a compact, easily accessible transmitter.
Accurate measurement of liquids and gases for a wide range of standard applications.
Sensor features Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Compact dual-tube sensor. Medium temperature up to +150 °C (+302 °F).
Transmitter features Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Process pressure up to 100 bar (1450 psi). Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access.
Nominal diameter range DN 8 to 80 (⅜ to 3")
Wetted materials Measuring tube: 1.4539 (904L)
Connection: 1.4404 (316/316L)
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
Max. measurement error Mass flow (liquid): ±0.15 % (standard), ±0.10 % (option)
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.50 %
Density (liquid): ±0.0005 g/cm³
Measuring range 0 to 180 000 kg/h (0 to 6615 lb/min)
Max. process pressure PN 100, Class 600, 63K
Medium temperature range –40 to +150 °C (–40 to +302 °F)
Ambient temperature range Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F)
Sensor housing material 1.4301 (304), corrosion resistant
Transmitter housing material AlSi10Mg, coated; stainless steel for hygenic transmitter design
Degree of protection Standard: IP66/67, Type 4X enclosure
IP69
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available"
Outputs 3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output
Inputs Status input
4-20 mA input
Digital communication HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, MODBUS TCP over Ethernet-APL, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Power supply DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvals ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
Product safety CE, C-tick, EAC marking
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Marine approvals and certificates LR approval, DNV approval, ABS approval, BV approval, CCS approval
Pressure approvals and certificates PED, CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, EHEDG, cGMP

Liquids

Measuring Principle Coriolis
Product headline Flowmeter with minimized total cost of ownership and a compact, easily accessible transmitter.
Accurate measurement of liquids and gases for a wide range of standard applications.
Sensor features Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Compact dual-tube sensor. Medium temperature up to +150 °C (+302 °F).
Transmitter features Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Process pressure up to 100 bar (1450 psi). Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access.
Nominal diameter range DN 8 to 80 (⅜ to 3")
Wetted materials Measuring tube: 1.4539 (904L)
Connection: 1.4404 (316/316L)
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
Max. measurement error Mass flow (liquid): ±0.15 % (standard), ±0.10 % (option)
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.50 %
Density (liquid): ±0.0005 g/cm³
Measuring range 0 to 180 000 kg/h (0 to 6615 lb/min)
Max. process pressure PN 100, Class 600, 63K
Medium temperature range –40 to +150 °C (–40 to +302 °F)
Ambient temperature range Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F)
Sensor housing material 1.4301 (304), corrosion resistant
Transmitter housing material AlSi10Mg, coated; stainless steel for hygenic transmitter design
Degree of protection Standard: IP66/67, Type 4X enclosure
IP69
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available"
Outputs 3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output
Inputs Status input
4-20 mA input
Digital communication HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, MODBUS TCP over Ethernet-APL, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Power supply DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvals ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
Product safety CE, C-tick, EAC marking
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Marine approvals and certificates LR approval, DNV approval, ABS approval, BV approval, CCS approval
Pressure approvals and certificates PED, CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, EHEDG, cGMP

Steam

Measuring Principle Coriolis
Product headline Flowmeter with minimized total cost of ownership and a compact, easily accessible transmitter.
Accurate measurement of liquids and gases for a wide range of standard applications.
Marine approvals and certificates LR approval, DNV approval, ABS approval, BV approval
市政污水处理

市政污水处理污水处理厂 · 质量流量计应用案例

项目背景

市政污水处理客户在污水处理厂环节,长期面临「出水水质监测精度不足、仪表易受污染导致数据漂移,影响达标排放」的困扰,对质量流量计的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的质量流量计,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应市政污水处理复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
食品饮料

食品饮料食品生产线 · 质量流量计应用案例

项目背景

食品饮料客户在食品生产线环节,长期面临「卫生级测量要求高,在线仪表需耐 CIP/SIP 清洗且精度稳定」的困扰,对质量流量计的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的质量流量计,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应食品饮料复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
化工制药

化工制药反应釜与储罐 · 质量流量计应用案例

项目背景

化工制药客户在反应釜与储罐环节,长期面临「介质腐蚀性强、工况复杂,仪表需耐腐蚀且防爆」的困扰,对质量流量计的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的质量流量计,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应化工制药复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
电力能源

电力能源电厂锅炉与汽机 · 质量流量计应用案例

项目背景

电力能源客户在电厂锅炉与汽机环节,长期面临「高温高压工况下测量不稳定,维护停机成本高」的困扰,对质量流量计的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的质量流量计,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应电力能源复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。