Instrumentation · Engineering · Package · PLC
0755-8222 1901 info@ainstru.com

CYZ71D E+H Multi Parameter Sensor Management Software

CYZ71D E+H Multi Parameter Sensor Management Software

多参数传感器管理软件

After installing the CYZ71D software on a computer tool, it is possible to measure, calibrate, and document Memosens sensors, which is suitable for laboratory measurements.
Compare

Conductivity

Measuring Principle Conductive
Application Measurement, calibration and documentation of Memosens sensors in laboratories, up to four channels
Installation Laboratory or work bench applications in safe area
No connection to PCS!
Characteristic Product consists of PC software, USB cable, sensor connection box MemoLink and Memosens lab cable CYK20 optionally, CYK10 with M12 plug can be used alternatively
Measuring range Dependent on connected sensors, please see sensor specifications
Measuring principle Memosens + measuring principles of the connected sensors
Material MemoLink sensor connection box: PBT
Dimension MemoLink sensor connection box:
97.0 x 31.5 x 77.5 mm
3.82 x 1.24 x 3.05 inch
Ex certification ATEX II (2)G (Ex ia Gb) IIc which enables to connect Ex- and non-ex sensors alternating to the MemoLoink box without loosing the sensor's approval
PC, USB cable and MemoLink in safe area only!
Ingress protection MemoLink: IP 65 when cables are connected
Input Memosens
Output / communication USB
Certificates, declarations and approvals FDA CFR 21 Part 11 Certificate of compliance when Pharma conformity is ordered

Oxygen

Measuring Principle Amperometric oxygen measurement
Application Measurement, calibration and documentation of Memosens sensors in laboratories, up to four channels
Installation Laboratory or work bench applications in safe area
No connection to PCS!
Characteristic Product consists of PC software, USB cable, sensor connection box MemoLink and Memosens lab cable CYK20 optionally, CYK10 with M12 plug can be used alternatively
Measuring range Dependent on connected sensors, please see sensor specifications
Measuring principle Memosens and measuring principles of the connected sensors
Material MemoLink sensor connection box: PBT
Dimension MemoLink sensor connection box:
97.0 x 31.5 x 77.5 mm
3.82 x 1.24 x 3.05 inch
Ex certification ATEX II (2)G (Ex ia Gb) IIc which enables to connect Ex- and non-ex sensors alternating to the MemoLoink box without loosing the sensor's approval.
PC, USB cable and MemoLink in safe area only!
Ingress protection MemoLink: IP 65 when cables are connected.
Input Memosens
Output / communication USB
Certificates, declarations and approvals FDA CFR 21 Part 11 Certificate of compliance when Pharma conformity is ordered

pH

Measuring Principle Potentiometric
Application Measurement, calibration and documentation of Memosens sensors in laboratories, up to four channels
Installation Laboratory or work bench applications in safe area, no connection to PCS!
Characteristic Product consists of PC software, USB cable, sensor connection box MemoLink and Memosens lab cable CYK20 optionally, CYK10 with M12 plug can be used alternatively
Measuring range Dependent on connected sensors, please see sensor specifications
Measuring principle Memosens and measuring principles of the connected sensors
Material MemoLink sensor connection box: PBT
Dimension MemoLink sensor connection box:
97.0 x 31.5 x 77.5 mm
3.82 x 1.24 x 3.05 inch
Ex certification ATEX II (2)G (Ex ia Gb) IIc which enables to connect Ex- and non-ex sensors alternating to the MemoLoink box without loosing the sensor's approval
PC, USB cable and MemoLink in safe area only!
Ingress protection MemoLink: IP 65 when cables are connected
Input Memosens
Output / communication USB
Certificates, declarations and approvals FDA CFR 21 Part 11 Certificate of compliance when Pharma conformity is ordered

Memobase Plus is the perfect software to improve the safety and reliability of your measurement process. Its documentation function can provide complete traceability documents for the entire life cycle of the Memosens sensor with the most streamlined documentation. Memobase Plus software turns your computer into a small, high-performance measuring device that can connect up to four sensors in parallel. The use of sensors in the process field is no different from that in the laboratory, with comparable measurement results.

 

advantage

high-precision sensor, low maintenance workload:

Plug-and-play Memosens sensors mean that process measurements can be performed simply by replacing sensors pre-calibrated in the laboratory

advanced diagnostic function and "readiness" display: Memobase Plus evaluates the sensor performance and clearly displays the sensor status through visualization to prevent non-compliant sensors from entering the measurement process and ensure process reliability

sensors are traceable for greater process safety:

Memosens sensors facilitate GLP, GMP, user management, and audit trails throughout the lifecycle, meeting FDA 21 CFR Part 11 standards

multi-channel, multi-parameter brings complete flexibility:

Memobase Plus transforms your PC into a space-saving, high-performance instrument with up to four Memosens sensors simultaneously

100% agreement between laboratory measurements and process measurements

Memobase Plus as a measurement workstation reduces the risk of discrepancies between laboratory and process measurements of samples

high precision measurement:

Memobase Plus is the tool of choice for true multi-point sensor calibration and calibration. Accurate measurements benefit from enhanced computational and statistical capabilities

simple buffer management:

inputs important information such as LOT number and validity period of buffer solution into Memobase Plus through scanning discussion-simple, fast and error-free, and convenient to track the type of buffer solution used in sensor calibration.

 

Application areas

Memobase Plus is suitable for any industrial field and fully meets the most demanding requirements of the pharmaceutical industry. User Management and Auditing Guidelines Ensure Measurement Process Compliance with FDA 21 Certification CFR, Part 11

 

Memobase Plus uses robust Memosens technology to manage the entire life cycle of pH, ORP, conductivity, dissolved oxygen and residual chlorine measurement sensors.

 

 

Technical Information

 

technical parameters

 

Documentation/Manuals/Software

 

Accessories/Spare Parts

Source:Endress Hauserwebsite public information

市政污水处理

市政污水处理污水处理厂 · 多参数应用案例

项目背景

市政污水处理客户在污水处理厂环节,长期面临「出水水质监测精度不足、仪表易受污染导致数据漂移,影响达标排放」的困扰,对多参数的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的多参数,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应市政污水处理复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
食品饮料

食品饮料食品生产线 · 多参数应用案例

项目背景

食品饮料客户在食品生产线环节,长期面临「卫生级测量要求高,在线仪表需耐 CIP/SIP 清洗且精度稳定」的困扰,对多参数的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的多参数,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应食品饮料复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
化工制药

化工制药反应釜与储罐 · 多参数应用案例

项目背景

化工制药客户在反应釜与储罐环节,长期面临「介质腐蚀性强、工况复杂,仪表需耐腐蚀且防爆」的困扰,对多参数的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的多参数,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应化工制药复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
电力能源

电力能源电厂锅炉与汽机 · 多参数应用案例

项目背景

电力能源客户在电厂锅炉与汽机环节,长期面临「高温高压工况下测量不稳定,维护停机成本高」的困扰,对多参数的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的多参数,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应电力能源复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。