Technical Data
ComparepH
| Measuring Principle | Potentiometric |
| Application | Hygienic and sterile applications (sterilizable, autoclavable): • Bioreactor/fermenter • Biotechnology • Pharmaceutical industry • Foods |
| Characteristic | Digital pH electrodes for hygienic production processes with ion trap for long-term stable reference |
| Measuring range | pH: 0 to 14 |
| Measuring principle | Gel compact electrode with ceramic junction and ion trap |
| Design | All shaft lengths with temperature sensor Advanced gel technology |
| Material | Sensor shaft: Glass to suit process pH membrane glass: Type N Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
| Dimension | Diameter: 12 mm (0.47 inch) Shaft lengths: 120, 225, 360 and 425 mm (4.72, 8.86, 14.2 and 16.7 inch) |
| Process temperature | Application N: 0 to 100 °C (32 to 212 °F) Up to 140 °C (284 °F) for sterilization |
| Process pressure | Application M: 0.8 to 14 bar (11.6 to 203 psi) absolute Application N: 0.8 to 7 bar (11.6 to 101.5 psi) absolute |
| Temperature sensor | NTC 30k |
| Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2. |
| Connection | Inductive, contactless connection head with Memosens 2.0 technology |
| Ingress protection | IP68 |
Field of application
Keep the pH value in the optimum range and maximize your product yield with Memosens CPS61E. Its rugged design and long-term stability ensure extremely accurate and reproducible measured values even after CIP/SIP or autoclaving. CPS61E features Memosens 2.0 digital technology, offering extended storage of calibration and process data for predictive maintenance. The sensor resists moisture and enables lab calibration, enhancing process integrity and increasing process uptime.
Memosens CPS61E measures pH in:
Biotech and pharma industries
- Bioreactors/fermenters: process control in enzyme production
- Bioreactors/fermenters: control of culture growth
Food & Beverage
- CIP/SIP applications
With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
A pharma certificate of compliance with sensor serial number specified is available.
Benefits
Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
The sensor's exceptional accuracy and reproducibility help you to keep the pH value in the optimum range for a maximized product yield.
A special glass membrane and reference system makes the sensor CIP/SIP and autoclaving resistant (up to 140°C/284°F) offering maximum long-term stability.
The sensor protects your product quality with its certified biocompatibility according to USP class VI, FDA compliance of wetted parts, VO 1935/2004 compliance, no cytotoxicity, bioreactivity or animal-based materials.
Streamline maintenance of CPS61E with the Memobase Plus software that automatically stores and documents all relevant sensor and process data.
Maximum process safety through non-contact, inductive signal transmission
Reduced operating costs thanks to minimized process downtime and extended sensor lifetime.
pH:
| Measuring Principle | Potentiometric | |
| Application | Hygienic and sterile applications (sterilizable, autoclavable): • Bioreactor/fermenter • Biotechnology • Pharmaceutical industry • Foods |
|
| Characteristic | Digital pH electrodes for hygienic production processes with ion trap for long-term stable reference | |
| Measuring range | pH: 0 to 14 | |
| Measuring principle | Gel compact electrode with ceramic junction and ion trap | |
| Design | All shaft lengths with temperature sensor Advanced gel technology |
|
| Material | Sensor shaft: Glass to suit process pH membrane glass: Type N Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
|
| Dimension | Diameter: 12 mm (0.47 inch) Shaft lengths: 120, 225, 360 and 425 mm (4.72, 8.86, 14.2 and 16.7 inch) |
|
| Process temperature | Application N: 0 to 100 °C (32 to 212 °F) Up to 140 °C (284 °F) for sterilization |
|
| Process pressure | 0.8 to 7 bar (11.6 to 101.5 psi) absolute | |
| Temperature sensor | NTC 30k | |
| Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2. | |
| Connection | Inductive, contactless connection head with Memosens 2.0 technology | |
| Ingress protection | IP68 | |
Related Models
| Model | Name |
|---|---|
| CPS61E-AA6MTU2 | pH电极 |
| CPS61E-AA6NTP2 | pH电极 |
| CPS61E-NA6NTP2 | pH电极 |
Case Studies
市政污水处理污水处理厂 · PH应用案例
项目背景
市政污水处理客户在污水处理厂环节,长期面临「出水水质监测精度不足、仪表易受污染导致数据漂移,影响达标排放」的困扰,对PH的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应市政污水处理复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
食品饮料食品生产线 · PH应用案例
项目背景
食品饮料客户在食品生产线环节,长期面临「卫生级测量要求高,在线仪表需耐 CIP/SIP 清洗且精度稳定」的困扰,对PH的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应食品饮料复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
化工制药反应釜与储罐 · PH应用案例
项目背景
化工制药客户在反应釜与储罐环节,长期面临「介质腐蚀性强、工况复杂,仪表需耐腐蚀且防爆」的困扰,对PH的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应化工制药复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
电力能源电厂锅炉与汽机 · PH应用案例
项目背景
电力能源客户在电厂锅炉与汽机环节,长期面临「高温高压工况下测量不稳定,维护停机成本高」的困扰,对PH的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应电力能源复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。